# 8. Bench and Close-Range Data

**Goal:** see that WISER treats a laboratory cube exactly like an airborne one,
and learn what a bench instrument needs from you that an airborne product has
already had done to it.

**Data:** `src/test_utils/test_datasets/circuit_4_100_150_um.hdr` — a 150 × 150
pixel, 4-band close-range scene of a printed circuit board at 0.525, 0.635,
0.740 and 1.200 µm. Note the units: micrometers, not nanometers.

**Reference:** {doc}`Opening Data Files <../user-content/opening-data-files>`,
{doc}`Spectra and Spectral Libraries <../user-content/spectra-and-libraries>`,
{doc}`Filters and Smoothing <../user-content/filters>`.

---

## A cube is a cube

Nothing in WISER assumes an aircraft or a satellite. A cube is *x*, *y* and
wavelength, and every tool in the previous seven tutorials works the same way on
one measured 30 cm from a bench as on one measured from orbit. Bench-scale
imaging spectrometers sort plastics for recycling, verify pharmaceutical
tablets, detect bruising in fruit, authenticate pigments in paintings, and
inspect electronics.

This tutorial opens a close-range scene, reads spectra off it, and then spends
most of its length on the part that actually differs: what you have to do
yourself before laboratory measurements mean anything.

---

## Step 1 — Open it, and check the units

1. Choose **File ▸ Open...** and select `circuit_4_100_150_um.hdr`.
2. If any of the four panes is not showing, click its pane toggle on the Main
   Toolbar. Then click **Zoom to fit**.
3. Click **Stretch builder**, select **Linear Stretch**, click **2.5% linear**,
   and click **OK**.

:::{figure} ../_static/tutorials/t8_board_rgb.png
:width: 90%
:align: center
:alt: The circuit-board scene open in WISER with all four panes visible, zoomed to 535 percent
:::

At 150 × 150 pixels over a few centimeters of board, individual pixels are
large on screen. The dark rectangles are component bodies, the pale field is
board substrate, and the small bright spots are pads and plated through-holes.

Look at the **Dataset Info** pane and find the wavelengths. They read 0.525 to
1.200 in micrometers, because that is what the ENVI header declares, and WISER
labels the spectrum-plot axis to match rather than assuming nanometers.

```{admonition} Interpretation
:class: note
Confirm the wavelength units on any instrument you have not used before. A cube
whose axis is off by a factor of a thousand will still plot, and will still be
wrong against every library you compare it against. Nothing in the software can
catch this for you, because both are valid units.
```

---

## Step 2 — Collect a spectrum per material

1. Click one pixel on a dark component body. Click **Collect spectrum** on the
   Spectrum Plot toolbar.
2. Right-click the new spectrum in the list below the plot, choose **Edit...**,
   and set **Spectrum name:** and **Plot color:** in the **Spectrum
   Information** dialog.
3. Repeat for a pixel on the pale board substrate and one on a bright pad.

Pick pixels in the middle of a uniform area. At this pixel size an edge pixel is
a mixture of both sides.

:::{figure} ../_static/tutorials/t8_board_spectra.png
:width: 100%
:align: center
:alt: Three spectra from the circuit board, well separated in brightness, with the component body nearly flat and the substrate falling steeply into the near infrared
:::

The component body is dark and almost flat. The board substrate falls
steadily across the range. The bright pad is highest everywhere and turns over
slightly, peaking at 0.635 µm rather than at the shortest band.

Most of what separates these curves is brightness, but not all of it. Divided
by its own mean, each curve has a different shape:

| | 0.525 | 0.635 | 0.740 | 1.200 µm |
|---|---|---|---|---|
| Component body | 1.14 | 0.94 | 0.96 | **0.96** |
| Board substrate | 1.18 | 1.08 | 0.99 | **0.75** |
| Bright pad | 1.04 | 1.09 | 1.01 | **0.86** |

The last column is the one that carries information a grayscale photograph
would not: the substrate loses a quarter of its relative reflectance by
1.200 µm while the component body holds nearly flat. That difference, not the
brightness, is what a classifier should key on.

---

## What four bands cannot do

Be honest about the limits of this scene before drawing conclusions from it.
The four bands are strongly correlated (the three visible ones at 0.986 to
0.995), and a principal-component transform puts **97% of the variance in the
first component**, with all four loadings the same sign. A first component like
that is brightness. Run K-means on this cube and the clusters come back sorted
by brightness, which is what a threshold on a single band would have given you.

So this fixture is the right size to practice the mechanics on and the wrong
size to identify a material with. Naming the three curves above took looking at
the image, not looking at the spectra. Identification needs narrow, diagnostic
absorptions, and those need tens to hundreds of bands: the C–H overtones that
separate plastics sit near 1700 and 2300 nm, and nothing in this cube reaches
past 1200 nm.

That is the argument for a real instrument rather than a four-filter camera,
and the rest of this page is about using one.

---

## Taking this to your own instrument

Your own bench system will give you hundreds of bands. What to get right:

**Calibrate to reflectance.** Image a white reference panel (Spectralon or
similar) and a dark frame under the same illumination and exposure, then
compute

$$R = \frac{S - D}{W - D}$$

with band math. Set **Type** to **Image Cube** for all three variables: WISER
guesses a type from each name, and it guesses **Spectrum** for `S` and **Image
Band** for `W` and `D`. Everything spectroscopic depends on this calibration.
Raw digital numbers carry your lamp's spectrum and your sensor's response, and
no reference library will match them.

**Watch the geometry.** Close-range imaging has strong, uneven illumination and
specular highlights. A shiny solder pad can saturate at one angle and read
near-zero at another. Use diffuse illumination, and prefer
{doc}`SAM <07-detection>` and continuum removal, both of which discount
brightness, over methods that depend on absolute level.

**Save your own library.** Image known reference materials, collect their
spectra, and export them as an ENVI spectral library. Then SAM and SFF work
against *your* materials under *your* optics, which is often better than a
published library for the specific question you are asking.

**Smooth before differentiating.** Bench spectra of dark materials are noisy.
Right-click the image and choose **Filters ▸ Savitzky-Golay Filter...**. The
dialog asks for a **Window Length (odd)** and a **Polynomial Order**. It smooths
along the spectral axis while preserving band shape and depth better than a
moving average does. The same menu offers **Mean**, **Median** and **Gaussian
Smoothing Filter...**, which can smooth spatially or spectrally. See
{doc}`Filters <../user-content/filters>`.

---

## Questions to think about

1. Why must a bench cube be converted to reflectance before you compare it with
   a spectral library?
2. A specular highlight saturates a copper pad in three bands. What does that
   do to a SAM result for that pixel, and to a linear-unmixing result?
3. You want to distinguish two visually identical black plastics. What would
   you need from your instrument that a four-band system cannot give you?
4. Why is a library you measured yourself often better than a published one for
   an inspection task?

---

## Where this is used

| Field | Scene | What to look for |
|---|---|---|
| **Recycling** | Mixed plastic flakes | C–H overtones, 1600–1800 nm and 2200–2400 nm; separates PET / HDPE / PP / PS |
| **Food quality** | Fruit surface | Water at 970 nm and 1450 nm; bruising shows before it is visible |
| **Pharmaceutical** | Tablets | API and excipient distribution; blend uniformity |
| **Cultural heritage** | Painting or manuscript | Pigment identification; underdrawing in the NIR; retouching |
| **Forensics** | Documents, fibers | Ink discrimination where inks are visually identical |
| **Soil science** | Core or sample tray | Organic carbon, clay mineralogy, moisture |

---

## What you can now do

- Open a close-range cube and confirm its wavelength units before trusting them
- Read spectra off materials that are not landscape
- Tell brightness differences from shape differences, and say which your data
  can support
- Calibrate a bench cube to reflectance, and build a library for your own optics

---

**Next:** back to the {doc}`tutorials index <index>`, or on to the
{doc}`User Manual <../user-content/user-manual>` for the full reference.