Filters and Smoothing#
WISER has four filters for suppressing noise in a cube. They are on the raster view’s context menu, under Filters: right-click the image and choose the one you want. There is no Tools-menu equivalent.
Filter |
Smooths along |
Preserves |
|---|---|---|
Savitzky–Golay |
Wavelength |
Absorption band shape and depth |
Mean |
Your choice of axis |
Nothing in particular — fastest |
Median |
Your choice of axis |
Edges; removes isolated spikes |
Gaussian |
Your choice of axis |
Smooth structure; well-behaved weighting |
Each produces a new dataset; the input is untouched. Runs proceed in the background; track them in the Activity Monitor.
Filter for a reason
Smoothing throws information away. Filter when you have a specific noise problem, such as a spectrum you cannot read, salt-and-pepper speckle or striping, and not as a routine first step. In particular, do not smooth before MNF: MNF’s noise estimate assumes the noise it finds is the sensor’s, and pre-smoothing invalidates that.
Savitzky–Golay filter#
The Savitzky–Golay filter fits a low-order polynomial to a sliding window of each pixel’s spectrum by least squares and replaces the center value with the fitted one. Because it fits a curve rather than averaging, it removes noise while leaving the position, shape and depth of absorption bands close to intact. A moving average destroys exactly those.
This is the standard smoothing step in spectroscopy, and the right choice whenever you intend to measure a band afterwards.
Right-click the image and choose Filters ▸ Savitzky-Golay Filter….
Setting |
Meaning |
|---|---|
Window Length (odd) |
Number of bands in the fitting window. Must be odd, and greater than the polynomial order. Larger smooths more. |
Polynomial Order |
Degree of the fitted polynomial. 2 or 3 is usual. |
Choose Dataset |
The cube to filter. |
Choosing the parameters. Start with a window of 5–9 bands and order 2 or 3. The window should be narrower than the narrowest feature you care about: a 21-band window will flatten a 15-band absorption. Raising the order preserves more curvature at the cost of removing less noise.
The filter runs along the spectral axis only.
Mean, median and Gaussian smoothing#
These three share one dialog, reached by right-clicking the image and choosing Filters ▸ Mean / Median / Gaussian Smoothing Filter….
Setting |
Meaning |
|---|---|
Input Dataset |
The cube to filter |
Axis |
Spectral — along wavelength, per pixel. Spatial — within each band, across the image. |
Mode |
Edge handling: |
Constant Value (cval) |
The value used outside the edge when Mode is |
Sigma |
Gaussian only — standard deviation of the kernel |
Radius / Truncate |
Gaussian only — where to cut the kernel off, as a radius in samples or a multiple of sigma |
Size |
Mean and median only — the window width |
WISER remembers the parameters you typed for each axis separately, so switching between Spectral and Spatial does not lose what you entered for the other.
Which one#
Mean — cheapest. Blurs edges and is pulled around by outliers.
Median — removes isolated spikes and dead pixels without blurring boundaries. The right choice for salt-and-pepper noise, and for a spectrum with a few bad bands you have not flagged.
Gaussian — weights nearby samples more than distant ones, giving a smoother result than a mean filter at the same width and no ringing. The default choice for general spatial smoothing.
Spectral or spatial#
Spectral smoothing reduces noise in each pixel’s spectrum, which is what you want before reading a shallow absorption band. Prefer Savitzky–Golay here: it does the same job with far less damage to band shape.
Spatial smoothing trades resolution for signal-to-noise: each band’s image gets cleaner and every pixel’s spectrum becomes a mixture of its neighbors’. Reasonable on a coarse scene where you want regional patterns; destructive when you care about small targets.
Tip
Averaging in the Spectrum Plot — the Default area-average size setting, mean or median over an n × n box — gives you the effect of spatial smoothing on the spectra you are looking at, without producing a whole new dataset. Use it when you only want to read a spectrum more clearly.
See also#
Minimum Noise Fraction — separates signal from noise rather than blurring both
Tutorial 8 — Bench and Close-Range Data — where Savitzky–Golay earns its place